Chapter 4
|
Specifications and part numbers
Figure 4.3. Open‑top
liner
Description
Drain line
Table 4.3. Typical open‑top mixing with standard impeller sizes.
Left-
to-right
offset
50 L
0
drum
100 L
0
drum
200 L
0
≈ 2 1/8 in.
drum
300 L
1/2 in. left
≈ 2 1/2 in.
drum
* 5:1 turndown ratio can be achieved using the adjustable angle motor head
Thermo Scientific
4.2.2 Open-top liners for 50 L, 100 L, 200 L, and 300 L
drums
Tubing set (ID x OD x L)
12.7 mm (1/2 in.) x 19.1 mm (3/4 in.) x 1.83 m (6 ft.)

4.3 General mixing specifications

Motor angle
Front-
to-back
Lift height
adjustable
position
motor head
≈ 1 in.
≈ 3/16 in.
≈ 2 in.
≈ 7/8 in.
≈ 12 1/16 in.
≈ 15.74 in.
Tank liner
Drain line
Minimum
(fixed or
Motor
working
rpm
liquid
volume
models)
10°
350
10 L
10°
350
20 L
10°
350
40 L
10°
350
79 L
HyPerforma DS 300 Single‑Use Mixer User's Guide
Tank liner
without a
drain line
End treatment
Polycarbonate quick connect 12.7 mm
(1/2 in.) insert and cap
Turn-
Minimum
Maximum
down
hold-up
liquid fill
ratio
5:1
100 mL
50 L
5:1
100 mL
100 L
5:1
100 mL
200 L
4:1*
500 mL
300 L
|