15.4 Calibration Menu Description

CALIBRATION
SYSTEM NAME
DRIVE CONFIG
1141471-L-00
LCD DISPLAY
SYSTEM NAME
INVACARE_
DRIVE CONFIG
>>4P CWD M91 & SP
2P RWD
2P CWD
4P RWD
4P RWD HD
4P CWD TDX
4P CWD HD
GB RWD
GB RWD
Create the Name for the System's Programming Settings
Name will be displayed on the Top Right corner of the MK6i Display.
Use the Programmer Left & Right Arrow keys to position the Insertion
Point ("_").
Use the Programmer Up and Down Arrow keys to change the letter /
number
Blank Spaces Not allowed. Name will end at that point.
Switches motor outputs to match appropriate drive configuration. Select the
configuration to match the wheelchair.
The wheelchair will not perform as designed without the correct drive
confiuration selected and saved.
This setting MUST be changed and saved each time a CMPJ+, PSR, PSF
or Display is added or replaced.
GB RWD and GB CWD are available only on the TDX - SR. G-Trac
controller supports 4P CWD M91 & SP, 4P RWD, 2P-CWD, 2P-RWD and
TDX-SI
DESCRIPTION
Calibration Menu