Invacare® MK6i™ Electronics
LCD DISPLAY
MOM REVERSE
> MOM
LTCH
DIGITAL 3 SPEED
> 3 SPEEDS
1 SPEED
SLEEP MODE
> ON
OFF
STBY SEL
> ON
OFF
50
PERFORMANCE ADJUSTMENT DESCRIPTION
Allows Reverse to be set as either MOMENTARY or LATCHED.
Allows Reverse to be set as either MOMENTARY or LATCHED.
On sets reverse as momentary, Off sets reverse in Latched.
Helpful for some sip-n-puff users backing into spaces.
Allows either 1 or 3 driving speeds for Digital Controls in Momentary Mode (Previously
named "Momentary Mode Select")
1 SPEED provides only one forward speed.
3 SPEEDS provides 3 forward speeds, (1/3, 2/3, or full programmed speed), selected
with a mode (re-set) switch prior to driving.
"3 SPEEDS" provides speed selections using digital driver controls - without
changing drives.
Present only when digital driver control is selected under Input Type.
Allows the Wheelchair to enter an "Inactive (resting) Mode" mode after a set period of
time with no driver control activity. (Previously named StandBy Mode)
Used with drivers who cannot access the On/Off switch during periods of no activity.
When chair enters sleep mode, a mode switch (Reset) is required to return the
wheelchair to operating mode.
Disappears from the menu if Standby Select is On.
Sends the wheelchair into resting mode. Driver commands then SELECT next operating
function – BYPASSING the "Reset Switch". Once in Standby mode:
FORWARD command ALWAYS returns the wheelchair back to Drive Mode.
RIGHT Command = Remote Drive Select Mode (if turned on). Subsequent Left
command changes to next drive.
LEFT Command = ECU functions, then Powered Seating Functions (if turned on).
Disappears from the menu if Sleep Mode is On.