CALIBRATION
TILT CALIBRATE
RECLINE CALIBRATE
CM LEGS CALIBRATE
BACK ANGLE
START IN DRIVE
1141471-L-00
LCD DISPLAY
TILT CALIBRATE
MOVE DOWN
SET DOWN ANGLE
MOVE UP
SET UP ANGLE
BACK ANGLE
95
START IN DRIVE
>>LAST USED
DRIVE 1
DRIVE 2
DRIVE 3
DRIVE 4
DESCRIPTION
Calibrating Tilt, Recline or Center Mount Legs, requires
a Pitch-Angle Gauge.
>
For Custom Actuators Only.
__°
1.
Select MOVE DOWN.
>
2.
Use the down arrow key to tilt the system all the way.
__°
3.
Press the Select key.
4.
Select SET DOWN ANGLE.
5.
Measure the angle of the seat using the Pitch-Angle
Gauge.
6.
Use the arrow keys to set the SET DOWN ANGLE to
the gauge measurement.
7.
Repeat for MOVE UP and tilt the system all the way up.
The angle of the back relative to the seat.
The back angle plus the tilt angle determines the drive
lockout angle.
A value between 85° and 105°, typically set at 95°.
Only displayed with smart actuators on tilt only
systems.
Allows Setting the Drive Mode (1 through 4) the wheelchair
powers up into.
RETURN TO LAST USED allows the wheelchair to
power up into the drive it was in when last powered
down.
DRIVE 1 allows the wheelchair to ALWAYS return to
1 when turned on.
Calibration Menu