Invacare MK6i Field Reference Manual page 46

Electronics
Hide thumbs Also See for MK6i:
Table of Contents

Advertisement

LCD DISPLAY
DCI MONITORING
(SPJ+ ONLY)
IVC MANUAL SEATING
(OR "CONTINUOUS")
IVC POWER SEATING
(OR "LATCHING")
ACC 1 (2)
(SPJ+ ONLY)
OFF
ON
TRACTION
30%
LESS
| | | | | | |
JSTK THROW
MOVE JOYSTICK TO:
FORWARD_ REVERSE_
LEFT_ RIGHT_ AND
THEN NEUTRAL_
1141471-L-00
PERFORMANCE ADJUSTMENT DESCRIPTION
Determines the method Drive Lock Out status is monitored by the controller.
CONTINUOUS/IVC MANUAL SEATING
Status of the drive lockout switch is continuously monitored for change.
Use this setting for manual tilt or recline systems.
Recommended for some aftermarket powered seating systems. Consult the
aftermarket seating manufacturer for more information.
LATCHING/IVC POWER SEATING
Monitors the Drive Lockout Input switch ONLY during power up, and after
actuator operation.
This setting is recommended for Invacare powered seating systems.
Allows setting actuator to operate through the SPJ+ with ACC Joysticks when ACC is
used on the TDX Spree Power Wheelchair with power tilt.
ACC 1 = ELEVATE, ACC2 = TILT
A reduction of the speed when going into and coming out of turns.
The higher the value, the greater the reduction in speed.
Set at 0% for the majority of wheelchair users.
MORE
Increasing values may be helpful to soften veer correction in "latched" driving mode,
or to dampen veer correction speed for aggressive drivers.
Used to calibrate joystick throw.
Sets the point for reaching full speed in relation to joystick displacement.
Values DO NOT return to default settings unless Manually Re-Set.
Reduces joystick movement required to reach full speed.
MK6i Performance Adjustments