Invacare MK6i Field Reference Manual page 108

Electronics
Hide thumbs Also See for MK6i:
Table of Contents

Advertisement

FAULT LOG
E32
E28
E04
E20
Clear Fault Log
A professional memory card is required in the MK6i
Programmer to view error code and troubleshooting
information.
Erases fault log list from view on the hand held programmer. Erased
faults can still be viewed through MK6 IVS.
CLEAR FAULT LOG
E32
E28
Tilt Actuator, Recline Actuator and CM Leg Actuator
The TILT ACTUATOR, RECLINE ACTUATOR AND CM LEG
ACTUATOR menu displays information about the actuators and
allows actuator movement.
Smart actuators in the system will show position angle and amp
draw.
Conventional actuators connected to the ACC controller will
not be displayed.
Conventional actuators connected using an actuator module
(e.g., TRAM (tilt recline actuator module) will show amp draw
not position angle.
POS - Displays the current position of the smart actuator.
Pressing the Up or Down arrow keys on the programmer will
move the actuator and display the position change and amp
draw of the actuator.
1141471-L-00
E09
E19
0
0
E09
E19
USE ARROWS TO MOVE
TILT
POS=__°
Version
The VERSION menu displays the software version of all components
recognized in the MK6i System.
VERSION
TRAM 1.33
4WSB 1.33
MPJ+ 1.3.0
User Settings Diagnostics
AMPS=±0.0
109

Hide quick links:

Advertisement

Table of Contents
loading