ERROR CODE
(None)
DEACTIVE displays
and the wheelchair will
not drive.
(None)
MAX BACK ANGLE
displays.
(None)
CONTROLLER
INHIBITED displays and
the wheelchair does not
drive.
(None)
SLOW DOWN
is displayed and the
wheelchair drives slowly.
1141471-L-00
SYMPTOM
PROBABLE CAUSE
If the wheelchair has a TIAM
and RIAM, this message is
displayed when the Tilt angle,
the recline angle or the
combined tilt and recline angle
are greater than 20° from
fully upright (beyond the drive
lock-out angle of 20°).
If the wheelchair has a TRAM
or TIAM, this message is
displayed when the tilt switch
is open.
The wheelchair back has
reached the maximum
programmed back angle
on a wheelchair with a TIAM,
RIAM or TRAM.
The system is tilted or reclined
beyond the drive lock-out
angle of 20°.
Incorrect ACC DCI setting in
the calibration menu.
The elevating seat is elevated.
SOLUTIONS
Return the system to upright position.
The wheelchair back will not go past the programmed
maximum back angle. This is normal behavior.
Return the system to the upright position.
Return elevating seat to the lowest position to drive at full
speed.