Invacare® MK6i™ Electronics
ERROR CODE
E41
CONTROLLER
STARTUP FAULT
displays and the
wheelchair drives slowly.
E102
GB GRNL FAULT
E103
GB FAULT - CYCLE
PWR
E104-E105
GB CTRL FAULT
E106-E107
GB CTRL FAULT
E108-E109
CURR CAL FAULT
E110-E111
GB CTRL FAULT
E112-E113
GB CTRL FAULT
E114-E127
CURR CAL FAULT
E128
M2 MOTOR FAULT
E129
M1 MOTOR FAULT
116
SYMPTOM
PROBABLE CAUSE
The controller has determined
a fault during a previous
turn-off process.
Unidentifiable Error
Possible Controller Failure
Left Current Sensor Error
Right Current Sensor Error
Current Calibration Error
Left (on CWD)/Right (on
RWD)
Current Sensor Error
Right (on RWD)/Left (on
CWD)
Current Sensor error
Current Calibration Error
Left Motor (M2) (on
RWD)/Right (on CWD)
Over Current Fault
Right Motor (M1) (on
RWD)/Left (on CWD) Over
Current Fault
SOLUTIONS
Turn the wheelchair off and back on.
Call Technical Services.
Turn chair off and then back on.
If fault repeats, replace controller and recalibrate motors.
Replace controller and recalibrate motors.
Replace controller and recalibrate motors.
No action required - Factory Test Only
Call Technical Services.
Replace controller and recalibrate motors.
Call Technical Services
Replace controller and recalibrate motors.
No action required - Factory Test Only
Check Left Motor (M2) (on RWD)/Right (on CWD) and
Cabling.
Check Right Motor (M1) (on RWD)/Left (on CWD) and
Cabling.
1141471-L-00