Invacare® MK6i™ Electronics
LCD DISPLAY
TORQUE
36
LESS
| | | | | | |
DCI operation (SPJ+ ONLY)
Off
AFTERMARKET
(OR "NORMALLY OPEN")
INVACARE
(OR "NORMALLY CLOSED")
46
PERFORMANCE ADJUSTMENT DESCRIPTION
A function of Time & Power. How quickly programmed Power Level is reached.
ALL FOUR quadrants are affected by the programmed torque level.
High = immediate ramping up to programmed power level. Recommended for Slow
MORE
Speeds, Switch drivers, rough terrains, curbs, significant obstacles.
Low = slower ramping up to programmed power level. Recommended for faster
speeds, level terrains, new drivers - even at some slower speeds.
MK6i Torque levels are noted in Ohms.
Small changes to torque have a significant impact.
Recommendation is to change only in 4 Ohm increments to determine if needs
are met.
Determines Drive Lock Out Function for systems utilizing any version of the SPJ+ Joystick
OFF
Disables Drive Lock Out for tilt or recline systems (manual or power),
Recommended setting for chairs without tilt or recline seating systems
NORMALLY OPEN/AFTERMARKET
Setting for some systems using aftermarket powered seating systems. Consult the
aftermarket seating manufacturer for more information.
NORMALLY CLOSED/INVACARE
Setting used for Invacare tilt or recline systems (manual or power).