Invacare MK6i Field Reference Manual page 100

Electronics
Hide thumbs Also See for MK6i:
Table of Contents

Advertisement

CALIBRATION
HARD PUFF
SOFT PUFF
HARD SIP
SOFT SIP
1141471-L-00
LCD DISPLAY
HARD PUFF CAL.
MIN
.70 MAX
S
| |
DESCRIPTION
Calibrates pressures required to activate hard / soft, puff
& sip of commands.
Separate screens for each of the 4 pressures.
H
Select command to calibrate. Follow instructions
below. Save on completion.
GOAL: Separate "S" & "H" values sufficiently for
easy distinction between Hard & Soft commands –
AND set values low enough to assure they can be
consistently achieved. Values should be between
.10 and 1.28.
In a Hard calibration mode, use the up/down arrows
to raise and lower the H value which MUST be met
as the user puffs/sips.
In a Soft calibration mode, use the up/down arrow
keys to raise and lower the S value, which MUST be
met as the user puffs/sips.
Calibration Menu
101