Verifying The Optical Connections And Optical Cw Operation - Keysight Technologies N4373D User Manual

Lightwave component analyzer
Hide thumbs Also See for N4373D:
Table of Contents

Advertisement

Performance Quick Check

Verifying the Optical Connections and Optical CW Operation

42
Passing these tests ensures that the transmitter and receiver are
within expected optical CW performance.
1. Follow the steps in the sections "Preparation " on page 38
and "Network analyzer settings and electrical calibration" on
page 38.
2. If you have option -021: 
Making sure the angled connector connects to the optical test
set, connect the two straight-angled optical patchcords
(N4373B-87907) to the optical output and input 1 ("+7 dBm")
or input 2 ("+17 dBm").
Use the FC/PC adapter (1005-0256) to connect the straight
ends of the optical patchcords together.
If you have option -022: 
Making sure the angled connector connects to the optical test
set, connect the angled-angled optical patchcords (N4373B-
87906) to the optical output and input 1 ("+7 dBm") or input
2 ("+17 dBm"). 
Use the FC/APC adapter (1005-1027) to connect the angled
ends of the optical patchcords together.
3. Press the [Macro/Local utility button until you see
4. Press
to start the LCA measurement setup.
5. Select the OO measurement tab.
6. Set the following parameters:
Keysight N4373D Lightwave Component Analyzer, User's Guide
.