• If you have option 102 (Dual wavelength 1310nm,
1550nm), set Wavelength to the wavelength at which you
are testing your DUTs (1310nm or 1550nm).
• Laser power: +4 dBm
• Optical input: Input 1
• User Calibration: None
• Forward RF Power: -1 dBm
7. Click
, then
8 Press
to start optical power meter application.
9 Read the optical power from the power meter application.
The readout must be +4 dBm ± 0. 5dB
10.Press the [Macro/Local utility button until you see
11.Press
to start the LCA measurement setup.
12.Select the OO measurement tab.
13.Set the following parameters:
• If you have option 102 (Dual wavelength 1310nm,
1550nm), set Wavelength to the wavelength at which you
are testing your DUTs (1310nm or 1550nm).
• Laser power: -1 dBm
Performance Quick Check
to set the output power.
.