Making Measurements
Not using "Continuous" modulator bias optimization affects
the accuracy of the internal transmitter calibration.
We recommend you include a "User Calibration" in your
setup ("User Calibration" on page 75) if you are not using
"Continuous" modulator bias optimization.
The system is calibrated with a default setting. You can
determine the default setting by going to the tools page,
choosing "Default", and pressing "Run".
7 If you have optical components, such as attenuators, fiber
optic cables or splitters, between the laser output and your
device under test, open and enable the Optical Path
Deembedding.
Here you can enter the parameters for the optical path
directly. 
Keysight N4373D Lightwave Component Analyzer, User's Guide