Updating Firmware

Important! As a precaution, you should download all tag IDs and short status reports contained
in memory before initiating the update process. Use the MED command to initiate the memory
download. All settings may be reset to manufacturer default values during the update process, so
it is recommended that you take a note of the present settings prior to updating the IS1001 reader.
New or updated firmware may periodically become available for IS1001 reader. To update the
firmware:
1. Start the BioTerm communication program.
Note: BioStat program does not support device firmware update.
2. Make a connection to the IS1001 reader using any available communication port.
3. Select Tools > Update Device Firmware within the BioTerm utility.
4. Go to the location of the new firmware file, select it and then click on the Open button.
Make sure you use the appropriate file for your IS1001 model:
IS1001: This model uses IS1001 Application Firmware v1.x.x.
IS1001-12V: This model uses IS1001 Application Firmware v2.x.x.
5. Click on the Start Programming button.