RST
{1,0}
RSV
{VV,VV}
RTS
{HH:MM:SS}
Reports
RDD
RDT
RFS
RFV
RND
{0-1440}
RNS
IS1001 Reader User Manual
Set Reader As Source For Date/Time Synchronization {1} On, {0}
Off
Enables real-time clock synchronization and sets the IS1001 reader
as the source. Valid values are On or Off. Default is Off.
Set Reader Automatic Standby Activation and Deactivation
Voltages (16-24,17-25)
Specifies at what input voltage IS1001 reader will automatically
switch into Standby mode to extend battery life and at what input
voltage it will switch back into Scan mode. Valid values are 16 to 24
V for activation of the feature (switching to Standby) and 17 to 25 V
for deactivation (switching back to Scan). Default values are 16 V
and 18 V respectively.
Set Reader Time (24-hour)
Sets the IS1001 reader's real-time clock present time. The time is
specified as hh:mm:ss (hours:minutes:seconds) in 24-hour format.
Report Diagnostic Data
Reports the IS1001 reader's vital diagnostics information.
Note: The data will be sent only to the port from which the report
was requested.
Report Reader Date and Time
Reports the IS1001 reader's real-time clock present date and time.
Note: The data will be sent only to the port from which the report
was requested.
Report Full Status
Immediately generates and sends a Full Status Report. This report
contains vital configuration settings and diagnostic data.
Note: The data will be sent only to the port from which the report
was requested.
Report Firmware Version
Immediately generates and sends a Firmware Version Report.
Note: The data will be sent only to the port from which the report
was requested.
Set Automatic Noise Report Delay in Minutes
Specifies how often to automatically send the Noise Report. Valid
values are 0 (disabled), 1 – 1440 minutes. Default is 0 (disabled).
Report Noise Status
Immediately generates and sends a Noise Report. The report
provides noise statistics such as present, average, and peak noise
Commands