Biomark IS1001 Operation User's Manual page 36

Standalone
Hide thumbs Also See for IS1001:
Table of Contents

Advertisement

Commands
DHT
{1|0}
DTD
{0-1440}
DTS
{1|0}
DTT
{1|0}
DUD
{1-43200}
DUM
{5|1|0|D}
32
Sets the FDX cycle time when HDX is enabled. Valid values are 45
– 500 milliseconds. Default is 120 milliseconds.
Set HDX Tag Detection {1} On, {0} Off
Specifies if HDX (half-duplex) tag detection is enabled. Valid values
are On or Off. Default is Off.
Set Automatic Virtual Test Tag Single Shot Delay in Minutes
Used to set the automatic Virtual Test Tag single shot activation
delay. The delay can be set from 0 (disabled) to 1440 minutes (24
hours). When the delay expires the Virtual Test Tag will be activated
for up to 120 milliseconds, unless the IS1001 reader has detected
the presence of a real PIT tag at its antenna, in which case the
Virtual Tag Test will be aborted and skipped. This test is done in
order to verify that the system is operating properly especially during
extended periods when no PIT tags are present.
Note: Virtual Test Tag will be activated for the amount of time equal
to the FDXB Detection Scan Time setting, but no longer than 120
milliseconds.
Set FDXB Tag Signal Level Detection {1} On, {0} Off
Enables FDXB tag detection signal level measurement. When
enabled, every time a FDXB tag ID is successfully detected, signal
level is measured and reported. The measured signal level value is
appended to the end of the detected tag ID message; it is also
stored with the tag ID in memory.
Valid values are On or Off. Default is Off.
Set BioTherm Tag Temperature Detection {1} On, {0} Off
Implemented BioTherm tag temperature detection. The measured
temperature value is appended to the end of the detected tag ID
message; it is also stored with the tag ID in memory. The displayed
temperature range is +25.0°C to +50.0°C. Temperature below
range is displayed as "LL.L C", temperature above range is
displayed as "HH.H C".
Valid values are On or Off. Default is Off.
Note: Specified BioTherm tag temperature measurement accuracy
is ±0.5°C, BUT only from +33.0°C to +43.0°C. Accuracy is not
specified and not guaranteed outside the +33.0°C to +43.0°C range.
Set Detection Unique Delay in Seconds
Specifies the delay value for the Detection Unique Mode Delay
option. Valid values are 1 – 43200 seconds. Default is 60 seconds.
Note: In order for Detection Unique Delay to take effect, Detection
Unique Mode must be set to Delay.
Set Detection Unique Mode {5} Last 5, {1} Last 1, {0} Off, {D} Delay
Specifies how the tag ID is handled if it is detected multiple times by
the same IS1001 reader. Valid values are:
IS1001 Reader User Manual