Alarm Messages and Codes

Alarm Messages and Codes
When an abnormal condition is detected by the IS1001 reader it generates an alarm message
that is sent to the communication ports. The alarms are sent to all available ports (unless the
CML0 command has been issued, in which case the alarms will not be sent to the local USB port.)
The following table is a list of the available alarms and their corresponding codes.
Code
Message
1
Reader Date/Time Lost
2
Reserved for MTS Mode
3
Reader Memory Error
4
Antenna Out Of Tune, Decrease
Capacitance
5
Antenna Out Of Tune, Increase
Capacitance
6
VTT Test Failed
7
Antenna Current Low
8
Noise High
9
Tuning Capacitance Low
10
Tuning Capacitance High
11
Input Voltage Low
46
Description
IS1001 reader date/time setting has been lost.
Set reader real-time clock present date and time.
IS1001 reader internal data memory reading/writing error detected.
Memory content is not guaranteed to be correct.
Unit requires repair.
The IS1001 reader is unable to reach the desired antenna tuning
using the internal capacitors bank.
Manual antenna capacitance reduction is required.
IS1001 reader is unable to reach the desired antenna tuning using the
internal capacitors bank.
Manual antenna capacitance increase is required.
IS1001 reader failed to detect VTT.
Verify antenna tuning and system noise level. VTT level adjustment
may be needed.
IS1001 reader has detected that the antenna current is below the
user-defined specification.
Verify antenna tuning and the reader's DC power supply. Exciter
voltage level adjustment may be needed. Alarm threshold adjustment
may be needed.
IS1001 reader has detected that the noise level is above the user-
defined specification.
Verfy antenna tuning. Verify surrounding environment for the cause of
the noise. Alarm threshold adjustment may be needed.
IS1001 reader has detected that the internal capacitors setting is
below the user-defined alarm threshold.
Verify antenna environment is not the cause of the problem (water
level significant change, metal objects in the antenna field, etc.).
Manual antenna capacitance adjustment may be needed. Tuning
capacitance low threshold adjustment may be needed.
IS1001 reader has detected that the internal capacitors setting is
above the user-defined alarm threshold.
Verify antenna environment is not the cause of the problem (water
level significant change, metal objects in the antenna field, etc.).
Manual antenna capacitance adjustment may be needed. Tuning
capacitance high threshold adjustment may be needed.
IS1001 reader has detected that the input voltage is below the factory
specifications. Alarm threshold depends on exciter voltage level
setting: 23 V for VE=5; 21 V for VE=4; 19 V for VE=3 and 18 V for
VE=1 and 2.
For IS1001-12V the threshold is 11.5 V.
Verify the reader's DC power supply.