5. Make the antenna connection to the IS1001 reader.
Connect the antenna leads to pins 1 (+) and 4 (-) of the 4-pin "ANTENNA" connector J8
on the Exciter Board.
Note: If there is a tuning capacitance connected in series at the antenna, the
recommendation is to connect the lead attached to it to Pin 4 and the lead attached to the
coil of the antenna to Pin 1; otherwise either lead can be connected to either of those two
pins. Both configurations should be tested to determine the one that yields the best
performance.
Connect the antenna cable shield drain wire to Pin 3 (GNDM) or to Pin 2 (GND) of the 4-
pin "ANTENNA" connector J8 on the Exciter Board, or leave it unconnected. You can
determine which configuration results in the best performance.
Adjust tuning capacitance range, if needed.