Information For Regular Tests; Main Components - sauter KERN SW 1000 Instruction Manual

Sound level meter
Hide thumbs Also See for KERN SW 1000:
Table of Contents

Advertisement

RTC
Language
Firmware update
Conditions
Real-time
temperature
Size (mm)
Weight
GPS
Calibrator
Printer
Note 1: For BSWA 200 instrument, ignore the result above 12.5 kHz due to the
microphone frequency response for Class 2 instruments.
Note 2: Data was measured with a microphone with a sensitivity of 50 mV/Pa for SW
1000 and 40 mV/Pa for BSWA 200.
Note 3: Measurement according to the requirements of GB/T3785 and IEC61672.

1.6 Information for regular tests

⚫ Reference sound pressure: 94,0 dB.
⚫ Reference direction of incidence : parallel to the direction of incidence of the
microphone.
⚫ Reference point of the microphone Center of the microphone diaphragm.
⚫ Reference direction of incidence: direction perpendicular to the microphone
diaphragm

1.7 Main components

Component name
Microphone
SW-BA-e-2020
also be provided by an external DC power source (7 V - 14
V, 500 mA) and the USB port (5 V, 1 A).
The integrated buffer battery has been calibrated at the
factory for an error of < 26 s in 30 days (< 10 ppm, (25 ± 16)
°C). The backup battery ensures that the system clock
continues to run even if the main batteries are changed.
GPS time function available (option with GPS module)
English, Chinese, Portuguese, Spanish, German, French
Firmware update via USB connection
Temperature: -10 ºC to 50 ºC. Humidity: 20 % - 90 %
Real-time temperature display on the main screen
Approx. 620 g with 4 alkaline batteries
Option
Receiver type: 50 channels Time to first measurement: Cold
start 27 s, warm start: 27 s, hot start 1 s; Sensitivity:
Tracking -161 dBm, new acquisition -160 dBm, cold start -
147 dBm, hot start -156 dBm; Horizontal position accuracy:
2.5 m, time accuracy: 30 ns, speed accuracy: 0.1 m/s;
update rate: 1 Hz, operating limits Dynamic ≤ 4 g, height <
50,000 m, speed < 500 m/s
CA111, Class 1, 94 dB/114 dB, 1 kHz
Mini thermal printer, RS-232 connection
Type
SW-A01
Class 1 microphone
SW-A02
Class 2 microphone
relative humidity
W 70 x H 300 x D 36
Description
7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sw 2000

Table of Contents