Measurement - sauter KERN SW 1000 Instruction Manual

Sound level meter
Hide thumbs Also See for KERN SW 1000:
Table of Contents

Advertisement

Select Calibration from the menu and press <Enter> to enter the By
3.
Measurement menu.
4.
Set the value for Cal.Level in the menu, for example to 93.8 dB. Calibration is
performed 5 seconds after pressing the <Start> button.
When the calibration is finished, the sound pressure meter updates the
5.
calibration factor. Press the <Enter> button to accept the results.
Go back to the main screen and press <Start/Stopp> to start the measurement.
6.
The current measurement result in this example is 93.8 dB if the calibrator is in good
working order.

4.3 Measurement

22
There are 13 menu options in the Measurement
menu. Use the arrow keys <▲>, <▼> to select
options and press <Enter> to move to the next
menu.
SW-BA-e-2020