Functional Upgrades; Specification - sauter KERN SW 1000 Instruction Manual

Sound level meter
Hide thumbs Also See for KERN SW 1000:
Table of Contents

Advertisement

⚫ Remote control connection RS-232
⚫ Mini thermal printer for printing the measured data
⚫ Internal GPS module (option), support of GPS time

1.4 Functional upgrades

Single chip high speed ARM with
floating point unit
LCD with white backlight
Integral period from 1 s to 24 h
New protocol steps 0.1 s, 0.2 s
and 0.5 s.
5 Templates for saving the user
settings
B weighting for ANSI standard
added
Automatic switch-on with external
power supply, easy integration

1.5 Specification

Type
Accuracy
Standard
octave1
Supplied
Microphone
Microphone interface
Detector / Filter
Integral period
Protocol step
Measurement
functions
SW-BA-e-2020
SW 1000
Class 1 (Group X)
GB/T 3785.1-2010, IEC 60651:1979, IEC 60804:2000
IEC 61672-1:2013, ANSI S1.4-1983, ANSI S1.43-1997
1/1 octave mid-frequencies: 31.5 Hz to 16 kHz
GB/T 3241-2010, IEC 61260-1:2014
ANSI S1.11-2004
MPA231T: Pre-polarized
measuring microphone 1/2",
class 1 Sensitivity: 50 mV/Pa.
Frequency range: 10 Hz to
20 kHz.
TNC connectors with ICCP supply (4 mA)
Purely digital signal processing with floating point unit
(digital detector and Filter)
User-definable integral period 1 s - 24 h [User].
Repeat time: infinite or 1 to 9999
0.1, 0.2 s, 0.5 s, 1 s to 24 h
LXY
, LXeq, LXYSD, LXSEL, LXE, LXYmax, LXYmin,
(SPL)
LXPeak, LXN. Where X is the frequency weighting: A, B, C,
Z; Y is the time weighting: F, S, I; N is the statistical portion:
1 to 99. 3 profiles and 14 user-defined measurements are
USB port implemented
Firmware update via USB (power
supply also via USB)
Support of automatic
measurements by timer function.
Internal GPS (option) with GPS
time
Only one measuring range to
cover the dynamic range of 123 dB
Reduced background noise (only
for class 1 instruments)
Upper measurement limit: 134
dBeff/137 dBeff (50 mV/Pa)
Class 2 (Group X)
1/1 octave center frequencies: 31.5 Hz to 8 kHz
GB/T 3241-2010, IEC 61260-1:2014
MPA200T: Pre-polarized
measuring microphone 1/2",
class 2 Sensitivity: 40 mV/Pa.
Frequency range: 20 Hz to
SW 2000
ANSI S1.11-2004
12.5 kHz.
5

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sw 2000

Table of Contents