4.3.12 24 h measurements with timer
The user can use the timer to implement a 24-hour measurement. The following
description uses an example to show how to implement the 24-hour measurement.
Activate SWN protocol and CSD protocol. Set the SWN protocol step to 1s and the
CSD protocol step to 5 min.
SW-BA-e-2020
Purpose: The measurement starts first on 14 March 2015 00:00 and
measures the first 5 min of each hour. The measured values are
stored in the CSD file when the measurement stops and every
second in the SWN file.
The delay setting in MEAS.setup is ignored if the measurement is
triggered by the timer. Set Itg.Period to 5 min and Repeat to 1.
Set the timer to loop mode so that the
measurement is triggered again and again.
Set the desired date as Start Day.
Set the Start Time as 00:00. This is the time at
which the measurement is triggered for the first
time.
Set the Repeat Interval to 1 h so that the
measurement is triggered every hour.

4.4 Setup

The setup menu contains the basic setup and
condition display functions. Use the arrow keys
<▲> , <▼> to select the option and press
<Enter> to move to the next menu level.
4.4.1 Contrast
The Contrast menu allows you to adjust the contrast of the LCD
display in 14 levels. Press the arrow keys <▲> , <▼> to select.