4.3.3 ICCP Power Supply
4.3.4 Profiles 1 to 3
Filter:
Define the filter for Profile 1 - 3 Press <◄>, <►> to select the following option: A,
B, C and Z (Z weighting means "no weighting". It is sometimes referred to as
"blanket" or "linear" weighting).
Detector:
Define the detector of profile 1 - 3 Press <◄>, <►> to select the following option:
Fast, Slow and Pulse.
mode:
Defines the integral mode of Profile 1 - 3 Press <◄>, <►> to select the following
option: SPL, PEAK, LEQ, MAX and MIN.
SWN Save:
With this option you define which data should be stored in the SWN file, since the
data source of the SWN file is Profile 1 - 3. This option has no relation to the screen
display. Press <◄>, <►> to select the following option: LEQ, PEAK, MAX or MIN.
4.3.5 Alarm threshold
4.3.6 Extended function
SW-BA-e-2020
The ICCP menu controls the power supply of all ICCP sensors via
the 24 V/4 mA constant current source. Disable the ICCP power
supply before connecting another sensor or connecting directly to
the signal source. Press the arrow keys <◄>, <►> to select.
The menu for the profiles 1 - 3 allows the definition of the filter, the
detector, the mode and the storage options of SWN files. Options
can be selected with the arrow keys <▲>, <▼>.
If measurement results of profile 1 - 3 exceed the alarm threshold,
the LED above <Power> will light red. The alarm threshold can be
set between 20 dB and 200 dB. Use the arrow keys <▲>, <▼> to
increase or decrease the alarm threshold by 1 dB. Use the arrow
keys <◄> , <►> to increase or decrease the setting by 10 dB.
With the advanced function, you can set which screen is displayed.
If the screen is not selected, it is not displayed. Note that the Main
Menu screen is always displayed.