3.2.1 CP924

1. Before starting connection work, hang the display unit of the CP924 in the extended bracket. (see red
arrows)
2. Connect the connections of the control unit according to the sections 4.2.1 und 4.2.2
3. Slide the holder into the enclosure and secure the display unit until it snaps into place.
4. When removing the display unit, make sure that you only place the suction cup in the area of the snap-
on devices and only carefully remove the display unit from the snap-on device.
CAUTION!

3.3 Digital inputs and outputs

CP9xx devices have 12 configurable digital inputs and outputs. The settings are made via the COMTRAXX® user
interface in a browser.
1
2
3
4
5
6
L
1
2
3
4
5
6
K

3.3.1 Functions

You can assign the following functions for the digital inputs:
off
Digital input without function.
TEST
Self test of the device.
RESET
Resetting error and alarm messages.
18
Damage to the display front
The front panel can be damaged by placing the suction siphon in the middle
and pulling it out at certain points..
7
8
9 10 11 12
L
7
8
9 10 11 12
K
Installation, connection and commissioning
Holder with retracted
slide
Snap mechanism
L12
L1
L2
...
N/O
N/C
K12
K1
K2
N/O
CP9xx_D00349_07_M_XXEN/01.2020