Installation, connection and commissioning

3.3.2 Mode

The operating mode for the digital input can be set to the following values:
•Active high
An event is carried out on the rising edge of the digital input (low to high).
Response time t(on)/t(off) after a switch-on signal
X1
X1
1
I
+
x
0
•Active low
An event is carried out on the falling edge of the digital input (high to low).
Response time t(on)/t(off) after a switch-off signal.
X1
X1
1
I
+
x
0

3.3.3 Response times t(on)/t(off)

The response time t(on) after a switch-on signal can be set between 100 ms and 300 s.
on
The response time t(off) after a switch-on signal can be set between 100 ms and 300 s.
off
3.4 Commissioning of the CP9... device
1. Switch on the supply voltage: After switching on, the device performs a start routine. It is completed
when the commissioning page appears n the display.
2. Enter the desired IP address in the 1st line
3. Enter the subnet mask in the 2
4. Enter the address of the default gateway.
5. Press the "Save" button to store the entries.
6. Wait 8-10 seconds.
7. If there is a DHCP server in the network, select only the check box to the right of the "DHCP?" label in
line 4. Confirm your selection by pressing the "Save" button. The network settings transmitted from the
server are displayed on the display after 8-10 seconds.