4.2.2 Memory scheme of the process image

4.2.2.1
Structure of the process image
As illustrated in the table, the Modbus start address for the respective process image is derived from the device
address. 256 (0x100) words or 512 bytes are reserved for each device. They contain all the information reques-
ted and transmitted from the bus.
Device address
1
2
3
...
32
...
255
Tab. 4.1: Modbus start address for each device for which a request can be sent.
4.2.2.2
Memory scheme of an individual device
Devices can feature various types of analogue and/or digital channels. Please take into consideration that there
are device-specific differences:
BMS devices usually feature 12 channels
MK800/TM800 supports up to 64 digital channels in the master mode
The channels 33 to 64 transmit digital messages only
Use the tables on Seite 23 and Seite 24 to determine the start address to request the following device parame-
ters:
Device type
Timestamp
Common alarm
Device error
BMS channel
CP9xx_D00349_07_M_XXEN/01.2020
Modbus address ranges of the process images in the memory
Word
HiByte
LoByte
...
00
0x01
Device 1
0x02
Device 2
0x03
Device 3
...
...
0x20
Device 32
...
...
0xFF
Device 255