Technical data

6. Technical data
Insulation coordination CP907 acc. to IEC 60664-1
Rated voltage.................................................................................................................................................................50 V
Overvoltage category ........................................................................................................................................................ III
Pollution degree ................................................................................................................................................................ 2
Rated impulse voltage.................................................................................................................................................800 V
Insulation coordination CP915 and CP924 acc. to IEC 60664-1
Rated insulation voltage........................................................................................................................................ AC 250 V
Overvoltage category ........................................................................................................................................................ III
Pollution degree ................................................................................................................................................................ 2
Rated impulse voltage................................................................................................................................................... 4 kV
Supply CP907 via plug-in terminal (A1/+;A2/-)
Nominal voltage CP907....................................................................................................................... DC 24 V SELV/PELV
Nominal voltage tolerance....................................................................................................................................+/-20 %
Typical power consumption at DC 24 V.................................................................................................................. < 15 W
Connection...........................................................................................................................plug-in terminal (A1/+;A2/-)
Maximum cable length when supplied via B95061210 (DC 24 V power supply unit 1.75 A):
0.28 mm²......................................................................................................................................................................75 m
0.5 mm²......................................................................................................................................................................130 m
0.75 mm²....................................................................................................................................................................200 m
1.5 mm²......................................................................................................................................................................400 m
2.5 mm²......................................................................................................................................................................650 m
USB power consumption ........................................................................................................................... DC 5 V, 500 mA
Supply via PoE (only CP907)
Nominal voltage.................................................................................................................................. DC 48 V SELV/PELV
Nominal voltage tolerance........................................................................................................................... -25...+15 %
Typical power consumption for PoE ....................................................................................................................... < 15 W
Maximum cable length when supplied via AWG 26/7; 0.14 mm² ..........................................................................100 m
Supply CP915/ CP924 via terminal block (L1; N)
Nominal voltage CP915 via external power supply unit......................................................................... AC 100... 240 V
Nominal voltage tolerance........................................................................................................................... -15...+10 %
Frequency range
......................................................................................................................................... 50...60 Hz
s
Typical power consumption at AC 230 V .............................................................................................. < 30 W / < 50 W
Connection........................................................................................................................................ terminal block (L1; N)
Stored energy time in the event of voltage failure
Time, date.......................................................................................................................................................... min. 3 days
Restart after voltage interruption...............................................................................................................min. 15 seconds
Displays, memory
Display CP907......................................................................................................................................7" TFT touch display
Display CP915.................................................................................................................................15.6" TFT touch display
Display CP924....................................................................................................................................24" TFT touch display
E-mail configuration and device failure monitoring................................................................................max. 250 entries
Individual texts ..........................................................................................................1200 texts with 100 characters each
Displayable devices ........................................................................................................................................................ 247
Number of data points for "third-party devices" to Modbus TCP and Modbus RTU ..................................................... 50
Number of data loggers ................................................................................................................................................... 30
Number of data points per data logger ....................................................................................................................10,000
Number of entries in the history memory ..................................................................................................................1,000
38