Bender COMTRAXX CP9 Series Manual page 17

Remote alarm indicator and operator panel for medical locations and other areas
Hide thumbs Also See for COMTRAXX CP9 Series:
Table of Contents

Advertisement

Installation, connection and commissioning
For CP915 only
9. Connect the front plate to the control board and the power supply unit:
Connect a USB port (9) to the board for the touch sensor connection on the front plate. Appropriate
connecting cable included in scope of delivery.
USB connection control board
9
10. Connect the screen output DVI (10) to the front plate. Appropriate DVI-D connecting cable included in
scope of delivery.
DVI-D-Anschluss Steuerplatine
10
11. Connect the voltage supply to the power supply unit via the pre-assembled wiring. Connect the
earthing to the front plate.
Spannungsversorgung vormontiert
12. Attach the front plate to the built-in flush-mounting box.
CP9xx_D00349_07_M_XXEN/01.2020
Touch sensor connection
Front plate