Modes; Safety Shutdown; Repair; Changing The High-Pressure Pump - Bosch DCI 700 Repair Instructions

Common-rail injector test bench
Hide thumbs Also See for DCI 700:
Table of Contents

Advertisement

1
3
1
2
1
 23:
Tightening torques
(1) Initial tightening 70 Nm; loosen 90°, second tightening 85 +5 Nm
(2) 110 +5 Nm
(3) 135 +5 Nm
 24:
Tightening torque for pressure line
(1) 60 +5 Nm
---
4.

Modes

4.1
---Separator---

Safety shutdown

The safety shutdown protects the DCI 700 from overheating and is activated in the following situations:
• Float switch B1 in the test oil tank has been actuated, e.g. in response to insufficient test oil in the test oil tank.
• Temperature switch B3 has been actuated at > 70 °C, e.g. in response to excessively high test oil temperature.
Resetting of the temperature switch takes place at < 55 °C.
• Temperature safeties B5, B6 or B7 in the heater have tripped, e.g. in response to lack of test oil in the test oil
tank.
Consequences
As soon as one of the above-mentioned safeties trips, the contactors Q2 and Q3 open. Contactors Q2 and Q3 are
no longer actuated, with the following consequences:
• Power is disconnected from the frequency converter A1 => Drive motor comes to a stop.
• The test oil heaters E1 – E3 are switched off => Test oil tempering ends.
• A fault message is displayed.
---Separator---
5.

Repair

---Separator---
5.1

Changing the high-pressure pump

5.1.1
---

Preparation

1. Close all applications and shut down the operating system.
2. Disconnect the DCI 700 at the main switch and secure against reactivation.
3. Unplug the mains plug or disconnect from mains electricity.
4. Remove attached injectors from the high-pressure rail. The swivel fixture with the high-pressure rail must swivel
freely.
---
Robert Bosch GmbH
1
1
1
1
DCI 700 | 19 | en
1 689 975 289 | 2020-05-06