Set The 8133A Pulse Width - Agilent Technologies 16900 Series Service Manual

Logic analyzer modules
Hide thumbs Also See for 16900 Series:
Table of Contents

Advertisement

1 On the 8133A pulse generator, set the Channel 2 pulse width to 1.5 ns.
2 Observe the 54845A oscilloscope display. Change the Channel 2 pulse
NOTE:

Set the 8133A pulse width

width of the 8133A pulse generator so that the pulse width measured at 1
volt on the oscilloscope is equal to 1.5 ns minus the measurement
uncertainty and display resolution of the oscilloscope, further reduced by
35 ps for test margin.
If you are using the 54845A/B oscilloscope, the measurement uncertainty
is ±((0.007% * ∆t) + (full scale/2x memory depth) + 30 ps) = ±30.10 ps.
Add 5 ps for display resolution. Add 35 ps test margin.
1.5 ns - 30.10 ps - 5 ps - 35 ps = 1.43 ns. Set the pulse width as measured
on the 54845A/B oscilloscope to 1.43 ns.
On the oscilloscope move the Ax and Bx markers to the crossing points of
the pulse and the horizontal center line. Read the pulse width at the
bottom of the screen. It is displayed as "∆=".
Chapter 3: Testing Logic Analyzer Performance
Connect the Test Equipment
39