Chapter 3: Testing Logic Analyzer Performance
Equipment Required
Equipment Required
The following equipment is required for the performance test procedure.
Equipment Required
Equipment
Pulse Generator
150 ps Transition Time Converter
(Qty 3)
Oscilloscope
SMA Coax Cable
(Qty 2)
Flying Lead Probe Set with 5
ground leads
(Qty 2)
Male BNC to Female SMA
adapters
(Qty 2)
SMA/Flying Lead test connectors,
(f) SMA to (f) SMA to Flying Lead
Probe
(Qty 3)
28
Critical Specification
≥ 260 MHz, two channels, differential
outputs, 150-180 ps rise/fall time (if faster,
use transition time converters)
Required if pulse generator's rise time is
less than 150 ps. (Pulse generator
conditions: Voffset=1V, ∆V=500 mV.)
Required for 8133A opt. 003.
bandwidth ≥ 1.5 GHz,
sampling rate ≥ 8 GSa/s
>18 GHz bandwidth
no substitute
no substitute
Recommended Model/Part
Agilent or HP 8133A option 003
Agilent or HP 15435A
Agilent or HP 54845A/B or similar
Agilent or HP 8120-4948
Agilent or HP E5383A
Cambridge Products CP-AD507
(see www.cambridgeproducts.com)
See "To Assemble the SMA/Flying Lead
Test Connectors" on page 22