To remove the logic analyzer cable
1 Remove power from the instrument
a In the session manager, select Shutdown.
b At the query, select Power Down.
c When the "OK to power down" message appears, turn the instrument
off.
d Disconnect the power cord.
2 Remove the logic analyzer cable.
a Use a T10 Torx driver to remove the screws that secure the logic
analyzer cable to the rear panel.
b Gently lift the logic analyzer cable end connector from the circuit board
connector.
3 If the logic analyzer cable is faulty, replace the cable and follow the next
procedure to install the replacement logic analyzer cable.
Chapter 6: Replacing Assemblies
85