Test Pods 3 and 4 in 500 Mb/s Mode (16910A only)

1 Perform the normal and complement tests for each additional pod on the
logic analyzer, changing the connection to the pod, channel assignments,
thresholds, etc. as appropriate. You must use clock Clk1 on Pod 1 for all
tests in the 500 Mb/s mode because the other clocks are not available in
this mode. Upon completion, the logic analyzer is completely tested.
2 Complete the Performance Test Record on page 60.

Conclude the State Mode Tests

Do the following steps to properly shut down the logic analyzer session after
completing the state mode tests.
1 End the test.
a In the Logic Analysis System window, select the [X] in the upper right
corner to close the window. At the query "Do you want to save the
current configuration?" select No.
Ending and restarting the logic analysis session will re-initialize the system.
b Disconnect all cables and adapters from the pulse generator and the
oscilloscope.
Chapter 3: Testing Logic Analyzer Performance
Test Pods 3 and 4 in 500 Mb/s Mode (16910A only)
59