Connect and configure the logic analyzer to test other pods
16800 Series Portable Logic Analyzers Service Guide
If the listing does not look similar to the figure, then
there is a possible problem with the cable or probe tip
assembly. Causes for cable test failures include:
•
Open channel.
•
Channel shorted to a neighboring channel.
•
Channel shorted to either ground or a supply voltage.
On the E5383A Flying Lead Probe set you have been
1
using, move Clk1 to the pulse generator Channel 2
OUTPUT (from OUTPUT).
Disconnect the logic analyzer Pod 1 cable from the flying
2
lead probe set and connect the logic analyzer Pod 2 cable.
Connect the second E5383A Flying Lead Probe set to the
3
logic analyzer Pod 1 cable.
4
Connect the second flying lead probe set's Clk1 to the 2 x
9 test fixture at the pulse generator Channel 2 OUTPUT.
In the Analyzer Setup dialog's Buses/Signals tab,
5
un- assign all channels of Pod 1.
6
Assign all channels of the pod to be tested. If the pod to
be tested has a clock, assign it too (place a check mark in
the box under the channel name).
Select the Sampling tab.
7
Click the Thresholds and Sampling Positions... button.
8
9
Move the starting position to - 2 ns.
5
Troubleshooting
111