Specifications

The specifications are the performance standards against which the product is
tested.
Individual
Data Channel
Parameter
Minimum master to
master clock time
tWidth (minimum)
Specifications verified under the following test conditions:
Parameter
Vh
Vl
vThreshold
rise/fall times
Probe
(Data Eye)
vHeight
Sampling
Position
(Eye Finder
Blue Bar)
Specifications
250 Mb/s
500 Mb/s
mode
mode
4 ns
2 ns
1.5 ns
1.5 ns
250 Mb/s
500 Mb/s
mode
mode
1.3 V
0.7 V
1 V
150-180 ps
Agilent E5383A
Chapter 1: General Information
tWidth
vThreshold
Notes
500 Mb/s mode is available only when
Option 500 is installed.
Specified at probe tip. Eye width as
measured by Eye Finder may be less.
Notes
600 mVp-p
Flying Lead Probe
0V
16753b01.vsd