Protect from moisture
Protection against penetration of solid foreign objects with a diameter greater than 12.5 mm, no
protection against water
Protection against penetration of solid foreign objects with a diameter greater than 12.5 mm, pro
tection against water dripping diagonally up to 15°.
Caution, hot surface
14.2 Operating states/error signals
The prosthesis indicates operating states and error messages through beeps and vibration signals.
14.2.1 Signals for operating states
Battery charger connected/disconnected
Beep signal
1 x short
1 x short
14.2.2 Warnings/error signals
Error during use
Beep signal
10 x long
Kenevo 3C60/3C60=ST
Vibration signal
None
3 x short
1x before beep signal
Vibration signal
1 x long at interval of
approx. 5 seconds
3 x long
5 x long
10 x long
Event
Battery charger connected or
battery charger already disconnected prior to start of char
ging mode
Charging mode started (3 sec. after connecting the bat
tery charger)
Battery charger disconnected after start of charging mode
Event
Overheated hydraulic unit Reduce activity.
Battery
charge
level
under 35%
Battery
charge
level
under 20%
Battery
charge
level
under 15%
After the beep and vibra
tion signals, the product
switches to empty battery
mode and then switches
off.
Appendices
Required action
Charge battery soon.
Charge battery immediately;
the product will be switched
off after the next warning sig
nal.
Charge the battery.
37