Connecting The Power Supply And Battery Charger; Connect Battery Charger To The Product - Otto Bock Kenevo 3C60 Instructions For Use Manual

Hide thumbs Also See for Kenevo 3C60:
Table of Contents

Advertisement

Charging the prosthesis battery

6.1 Connecting the power supply and battery charger

1
1) Slide the country-specific plug adapter onto the power supply until it locks into place (see fig. 1).
2) Connect the round, three-pin plug of the power supply to the receptacle on the inductive charger so that the
plug locks into place. (see fig. 2)
INFORMATION: Ensure correct polarity (guide lug). Do not use force when connecting the cable plug
to the battery charger.
3) Plug the power supply unit into the outlet (see fig. 3).
→ The green LED on the back of the power supply lights up.
→ The yellow LED on the inductive charger lights up briefly to indicate the correct connection to the power
supply.
→ If the green LED on the power supply does not light up and the yellow LED on the inductive charger does not
light up briefly while connecting the cable, there is an error (see Page 37).

6.2 Connect battery charger to the product

INFORMATION
Do not move the knee joint while it conducts the self-test immediately after disconnecting the charger. Otherwise,
an error may occur; if this happens, the problem can be corrected by reconnecting and then disconnecting the
charger.
INFORMATION
To make the operating time of the prosthesis as long as possible, the charger should not be removed until imme­
diately before the prosthesis is used.
18
2
1) Remove the prosthesis.
2) Connect the inductive charger to the receiver of the charging unit on the back of
the product.
Make sure the contact surfaces are clean, with no objects adhering to them.
→ The charger is held in place by a magnet.
→ A correct connection between the battery charger and the product is indic­
ated by feedback (see Page 38).
3) The charging process starts.
→ Once the product battery is fully charged, the LED on the battery charger
lights up green.
4) After the charging process is complete, remove the inductive charger from the
receiver and hold the product still.
→ A self-test is performed, and the product should not be moved while this is in
progress. The joint is ready for operation only after corresponding feedback
(see Page 38).
5) Put the prosthesis on.
3
Kenevo 3C60/3C60=ST

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Kenevo 3c60 st

Table of Contents