If the status light is blinking red, do not try to open the carousel. An AutoLoader error has occurred—refer to
`
"AutoLoader Troubleshooting" on page 50 for instructions.
4.
Remove a rack by slightly lifting up on the rack and pulling it out.
5.
Place slides into the rack with the slide labels facing towards the rear of the rack and the coverslips up.
Note – The slide racks can be loaded into any position on the AutoLoader. This is possible because the scanner
automatically senses which AutoLoader positions contain a slide rack. However you must use the Console to move
the rack position you want to use to load position before removing or replacing a rack. Do not try to manually rotate
the carousel.
Note that the rack has been designed to load slides from a Sakura Tissue-Tek automatic slide stainer and
coverslipper slide basket.
Aperio AT2 DX User's Guide, Revision B © Leica Biosystems Imaging, Inc. 2020
Chapter 6: Using the AutoLoader
37