Measuring - wtw inoLab Cond 730 Operating Manual

Laboratory conductivity meter
Table of Contents

Advertisement

Cond 730
Preparatory
activities
Temperature probe
4.2

Measuring

Perform the following activities when you want to measure:
1 Connect the measuring cell to the instrument.
2 Adjust the temperature of the test solutions or
measure the current temperature if the measurement
is made without a temperature probe.
3 Calibrate the instrument with the measuring cell or
check the cell constant set up.
4 Select the measuring mode by pressing
Caution
When connecting an earthed PC/printer, measurements
cannot be performed in earthed media as incorrect values
would result.
The RS232 interface is not galvanically isolated.
Measurements can be performed with and without a temper-
ature probe. A connected temperature probe is indicated by
TP on the display.
If you want to use a WTW conductivity measuring cell with-
out a temperature probe, you have to connect it with an
adapter (available at WTW).
Note
The conductivity meter automatically recognizes the type of
the temperature probe used. As a result, you can connect
electrodes with the NTC30 or Pt1000.
The temperature measurement is absolutely essential for a
reproducible conductivity measurement. If the measure-
ment is made without a temperature probe, proceed as fol-
lows:
1 Determine the current temperature using a
thermometer.
2 Set up the temperature by pressing
Operation
m
.
82
.
19