wtw inoLab Cond 730 Operating Manual page 70

Laboratory conductivity meter
Table of Contents

Advertisement

Technical data
Analog output
Serial interface
70
Automatic switch-over when the AK 323 recorder cable is
connected.
Signal range
0 ... 1,999 V for the range
0 ... 1999 digits
Accuracy
± 0,5 % of diplayed value
± 0,1 (% saturation)
or
± 0,5 %of displayed value ± 0.01 mg/l
Internal
< 5 ohms
resistance
(current limitation to max. 0.2 mA output
current)
Automatic switch-over when the AK 340/B or AK 325/S
cable is connected.
Type
RS 232, data output
Baud rate
can be set to 1200, 2400, 4800, 9600 Baud
Data bits
8
Stop bit
1
Parity
none
Handshake
RTS/CTS+Xon/Xoff
Cable length
max. 15m
Cond 730