Glossary
Monitors the electrode drift and releases the measured
value only after the stability criterion has been reached. In
this way, this procedure ensures the highest degree of pre-
cision and reproducibility.
The cell constant is determined through calibration. To do
so, the conductivity measuring cell is immersed into a series
of aqueous salt solutions with exactly known electric con-
ductivity. The relevant conductivity values are determined
using the conductivity measuring instrument.
Linear factor describing the geometrical dimensions of a
measuring cell and its electrodes.
l
With the value of the cell constant you can roughly descri-
be the application range of a conductivity measuring cell.
l
The cell constant is determined through calibration in one
or several calibration solutions.
The conductivity value is a sum parameter for the ion con-
centration of a test sample.
Solution with a known conductivity to determine or check the
conductivity.
See A
R
.
UTO
EAD
In order to compare values measured at different tempera-
tures, the values have to be converted to a fixed tempera-
ture. This temperature is 25 °C, or, as an exception, 20 °C.