wtw inoLab Cond 730 Operating Manual page 28

Laboratory conductivity meter
Table of Contents

Advertisement

Operation
Calibration interval
(Int 3)
Setting the
calibration interval
28
The flashing sensor symbol reminds you to calibrate regu-
larly. After the selected calibration interval (Int 3) expires,
the sensor symbol flashes. Measurements can continue.
Note
To ensure the high measuring precision of the measuring
system, perform a calibration after the calibration interval
expires.
The calibration interval (Int 3) is set to 180 days in the facto-
ry. The interval can be changed (1 ... 999 days):
1 Switch off the instrument.
m
2 Press
and hold down the key.
e
3 Press the
The display test appears briefly on the display.
The instrument then switches automatically to the
configuration level.
g
4 Press the
82
5 Press
next calibration.
6 Confirm with
7 Change to the measuring mode by pressing
key.
key until Int 3 appears on the display.
to set the required time interval until the
g
.
Cond 730
m
.