Keysight Technologies N9010B EXA Installation Note page 50

Mxa signal analyzer option b40, 40 mhz analysis bandwidth and microwave preselector bypass (mpb) upgrade kit
Hide thumbs Also See for N9010B EXA:
Table of Contents

Advertisement

Installation Procedure
Loading New Hardware Drivers
On the first boot-up after installing the new hardware, the analyzer will detect the new hardware
and will need to load drivers for the new hardware. The analyzer will boot-up as Instrument user,
but loading the new drivers requires administrator privileges.
1. Turn on the analyzer and wait for the XSA application to launch.
2. During the start up of the XSA application, it will detect the new hardware and need to load
new drivers. You will be prompted for an administrator login and password. The default
administrator login is "administrator" and the default password is "Keysight4u!".
3. You will receive two prompts as described above. The first prompt is for the "Keysight PXA DIF"
and the other is for the "Keysight Receiver Module". Both of these relate to the new A3 40 MHz
Digital IF Assembly.
4. After the hardware drivers have been installed, the XSA application will continue to start up.
Update FPGA Code
On the first boot-up after installing the new hardware, the analyzer will detect the new A3 40 MHz
BW Digital IF and A15 Front End Controller. However, the FPGA code in the new assemblies will not
necessarily be synchronized with the instrument software installed earlier. The following procedure
ensures that the FPGA code in the new assemblies is updated to the correct revision.
1. Turn on the analyzer and wait for the XSA application to launch.
2. Close the instrument application by pressing System and Exit Program. An Exit Program dialog
box will appear; tap OK to confirm the exit.
3. Tap Start, and select Log Off from the Shut Down drop-down list.
4. Tap the keyboard icon in the lower left or the Keyboard key below the numeric keypad.
Re-login as "administrator". The default password is "Keysight4u!". Tap OK.
5. On the desktop, tap Start, Computer, and select the C: drive. Navigate to the following folder:
C:\Program Files\Agilent\Signal Analysis\Physics
6. In this folder find and execute the file named:
FPGA_Prog.bat