Keysight Technologies N9010B EXA Installation Note page 20

Mxa signal analyzer option b40, 40 mhz analysis bandwidth and microwave preselector bypass (mpb) upgrade kit
Hide thumbs Also See for N9010B EXA:
Table of Contents

Advertisement

Installation Procedure
Rear Panel Removal
1. Refer to
Figure
8. Using the T-10 driver, remove the screws (2) attaching the rear panel (1) to
the chassis. The rear panel can now be removed.
2. Use a 9/16-inch socket wrench to remove the nut securing the EXT REF IN connector from the
rear panel.
Figure 8
Rear Panel Removal
3. Instruments with Option CR3 or CRP will have an additional cable W39 that is attached to the
rear panel. Remove W39 using either a 5/16" nut driver or a 5/16" open-end wrench.
4. The rear panel can now be removed.
20
Installation Note N9020-90253