Installation Procedure
2. Refer to
Figure
22. Pull the AIF/DIF assembly part way out of the chassis. Disconnect cables
W13, W14, and W15 from the AIF assembly.
Figure 22
AIF/DIF Assembly Removal
3. Discard cable W13.
4. The AIF/DIF assembly can be removed from the chassis by pulling straight out the back.
5. Separate the A2 Enhanced Analog IF and the A3 25 MHz Digital IF. The A2 will be re-used but
the A3 will be replaced with the A3 40 MHz Digital IF (N9020-60016) included in the kit. Retain
the screws.
6. Locate the new A3 40 MHz Digital IF (N9020-60016) included in this kit.
7. Connect the W2 ribbon cable to J20 on the DIF (2).