Installation Procedure
15.Refer to
Figure
28. Disconnect cable W12 from A16J710 on the A16 Reference Assembly and
discard.
Figure 28
Reference Cables
16.Refer to
Figure
28. Locate W54 in the Opt B40/BUN Cable Kit. This cable has part number
8121-2288 and will be labeled "716" on one end and "14" on the other. Connect the end
labeled "716" to A16J716 on the A16 Reference Assembly. Route the cable along the right side
wall to A3J14 on the A3 DIF Assembly. Add this cable to the spiral wrap around W14.
17.Refer to
Figure 24
have part number 8120-8868 and will be labeled "601" on one end and "19" on the other.
Connect the end labeled "19" to A3J19 on the A3 DIF Assembly. Route the cable to the right
side of the instrument and up to the A2 EAIF Assembly. Connect the cable to A2J601 with the
cable routed behind the screw head near A2J601.
18.Refer to
Figure
24. Connect W14 to A2J300 on the A2 EAIF Assembly.
40
and
Figure
25. Locate W41 in the Opt B40/BUN Cable Kit. This cable will
Installation Note N9020-90253