Keysight Technologies N9010B EXA Installation Note page 29

Mxa signal analyzer option b40, 40 mhz analysis bandwidth and microwave preselector bypass (mpb) upgrade kit
Hide thumbs Also See for N9010B EXA:
Table of Contents

Advertisement

14.Refer to
Figure
17. Attach the coaxial fixed attenuator (1) included in the kit to Switch 2
port 1. Torque to 10 inch-lbs.
Figure 17
Attenuator Installation
15.Locate W35 semi-rigid coax cable (N9020-20210) in the kit. Connect the cable between the
coaxial fixed attenuator (1) and port 1 of Switch 1 (the lower switch). Torque to
10 inch-pounds.
Installation Note N9020-90253
Installation Procedure
29