Installation Procedure
19.Refer to
Figure
20. Locate W33 semi-rigid coax cable (E4410-20165) in the kit. Connect the
cable between port 2 of Switch 1 (the lower switch) and the output of the A12 YTF Preselector.
Torque the cable at the A12 end first to 10 inch-pounds and then torque the cable at the
Switch 1 end to 10 inch-pounds.
20.Locate W32 semi-rigid coax cable (E4410-20163) in the kit. Connect the cable between port 2
of Switch 2 (the upper switch) and the input of the A12 YTF Preselector. Torque the cable at
the A12 end first to 10 inch-pounds and then torque the cable at the Switch 2 end to
10 inch-pounds.
Figure 20
Install W33 and W32
32
Installation Note N9020-90253