6. Refer to
Figure
15. Locate the other coaxial switch (N1810-60069) (1) and the PS Bypass
Switch 2 Bracket (E4410-00110) (2) in the kit.
Figure 15
Switch 2 Placement
7. Place the second switch/bracket assembly (1)/(2) on the bracket previously installed. Assure
Keysight label on switch is facing up. To attach use the two screws (3) (0515-1992) included
in the kit using the T-8 driver.
8. Secure the bracket to the Low Band Switch bracket with a single screw (4) (0515-0372)
included in the kit using the T-10 driver. Do not torque.
9. Align switches so they are parallel with the side of the instrument then torque the four screws
(3) and (5) securing the switches to the brackets to 6 inch-pounds using the T-8 driver.
Torque the screw (4) that secures the switch brackets together to 9 inch-pounds using the
T-10 driver.
Installation Note N9020-90253
Installation Procedure
27