Keysight Technologies N9010B EXA Installation Note page 23

Mxa signal analyzer option b40, 40 mhz analysis bandwidth and microwave preselector bypass (mpb) upgrade kit
Hide thumbs Also See for N9010B EXA:
Table of Contents

Advertisement

Replace Front End Control Assembly
1. Refer to
Figure
11. Remove the ribbon cables W16, W17, and W18 and the wire harness W22
from the Front End Control assembly (1).
2. The Front End Control assembly can now be unplugged from the motherboard by leveraging
up on the ejector and lifting the board up on the other side.
Figure 11
Front End Control Assembly Removal
Installation Note N9020-90253
Refer to Analyzer Information at the beginning of the Installation Procedure. Do not perform this
procedure if the Front End Controller has a Hw Id of 41 or 75.
Installation Procedure
23